Ideal for materials development
AC-2S Pro α
Features
・High light intensity irradiating optical system exceeding 2 μW
・Low-energy irradiating optical system capable of measurements from 2 eV
・High-temperature measurements up to 100 °C
・Film thickness measurements
・Precise measurements in 0.01 eV increments
・Includes often-requested multi-point measurement and repeated measurement functions
・Imports and analyzes AC-2/3/5 data
・Uses flat-plate open counter enabling photoelectron count rates of up to 4,000 cps
Applications
・Battery materials electron property measurements
・Catalyst materials electron property measurements
・Ionization potential measurements of organic light-emitting diodes and photoreceptors for photocopiers
・Tribology research on hard disks and magnetic tape
・Semiconductor and lead frame surface oxidization state measurements
・Detection of molecular level film contamination of precision electronic materials
・Ionization potential measurements of the following functional materials
Organic light-emitting diodes, organic solar cells, organic field-effect transistors, photoreceptor materials for photocopiers, Perovskite solar cells, quantum dots, carbon nanomaterials, catalyst materials, storage batteries, fuel cells, semiconductor materials (e.g., Si, GaN, GaAs), transparent conductive film (e.g., ITO, FTO)
Product category |
Photoemission Yield Spectroscopy in Air |
Model |
AC-2S Pro α |
Measurement principle |
Photoemission Yield Spectroscopy in Air (detector: low-energy photoelectron counting) |
Photoelectron detector |
Flat-plate open counter |
UV energy range |
2.0 - 6.2 eV (620 - 200 nm) |
Minimum light intensity |
5.0 nW or less (at 5.9 eV) |
Maximum light intensity |
2,500 nW or greater (at 5.9 eV) |
Repeatability (standard deviation) |
Work function 0.02 eV (sample: Au plate) |
UV light source |
Laser-driven light source (LDLS) |
UV spot size |
Up to 4 mm × 4 mm (varies according to energy due to chromatic aberration of converging lens) |
Maximum photoelectron count rate |
4,000 cps |
Spectrometer |
Grating type monochromator |
Maximum sample size |
Area up to 50 mm × 50 mm, thickness up to 10 mm |
Temperature/humidity range |
Temperature 15 °C - 35 °C (no sudden changes), up to 60 %RH (no condensation), dew point -30 °C or higher |
Power source |
Main unit: 100 V - 240 V AC 50/60 Hz, 5 A (max) |
External dimensions |
LC (light source unit)*: Approx. 623 mm (W) × 450 mm (D) × 317 mm (H) |
Weight |
AC-2S LC (light source unit): Approx. 30 kg, AC-2S DC (measurement unit): Approx. 31 kg |